Re: Chromosome sorting

John J. Fawcett (fawcett@lanl.gov)
Wed, 17 Jul 1996 09:31:51 -0600 (MDT)

>Is it possible to acheive good (chromosome) sorting when the first laser is 458
>and the second UV ?

Dr. Sharp,

Though I believe it is possible to obtain good chromosome
resolution when exciting with the 458 nm laser first, I would not
recommended it.

I prefer triggering on UV excited fluorescence since it is brighter
(usually 2-4X) and exhibits better chromosome peak resolution than
fluorescence that results from excitation at 458 nm.

=====================================================================
John J. Fawcett, fawcett@lanl.gov
Life Sciences Division, LS-3 505/667-9633 : office(vmail)
Los Alamos National Laboratory 505/667-2750 : LS Business Unit
P.O. Box 1663, ms M888 505/665-3024 : FAX
Los Alamos, NM 87545