Dear Bob,
Sorry to be so vague in my previous message.
>>What antigens do you assay for? Cytokines by sandwich immunoassay (avidin
coated beads; Capture: biotinylated monoclonal, Detection:Bodipy polyclonal)
The list: IL-1a,IL-1b,IL-1RA,IL-2,IL-6,IL-8,IL-12,IFNg and TNFa.
>>What quantity of antigen is on the beads (as/mg of beads)?
Luminex prepared the beads so they should comment on the bead loading
>>Are your results quantitative, or just +ve or -ve?
Quantitative; all the analyses produced curves with r2>0.99 with the
exception of IL-1b. IL-1b showed no binding to the beads at all (reagents?).
>>What is the assay detection limit for each antigen?
I didn't calculate them, but it appeared to be 2-3 fold higher than our kit
ELISA limits.

This experiment was a first attempt and not a definitive method validation
study; I do not want to draw too many conclusions from it. Kerry Oliver at
Luminex had only recently put this panel of cytokines together and had not run
anything but standards. However, we considered it enough of a success to keep
going.

Regards,
Gregg