Re: cell sorting

Marcus Reckermann (recker@ftz-west.uni-kiel.de)
Mon, 27 Oct 1997 11:42:05 +0000

> From: DoranC@AOL.COM
> Date: Thu, 23 Oct 1997 17:07:44 -0400 (EDT)
> To: Cytometry Mailing List <cytometry@flowcyt.cyto.purdue.edu>
> Subject: cell sorting

>
> I have recently been trained on the FACS Vantage and was told that while
> sorting cells, the abort rate should not be greater than 10% of the event
> rate (sys. threshold). My problem is that the abort rate is running approx.
> 20 - 30%. I acquire events usually around 2000, and most often sort
> normal-R, 3 drops. I have tried to alter the concentration, slow down the
> rate, and even filter the sample thru a filter cap. Any ideas?
>

What I have learned working with the Vantage ist that proper optical
alignment (i.e. y-knob, focussing the laser beam, and optimizing the
fluorescence sigmals) helps in reducing abort rates; also it is
crucial to have the dead time adjust knob on card cage slot 13
set properly. Play with that knob a little while running a
sample and monitoring the abort rate.
Good luck, marcus.
______________________________________________
Dr. Marcus Reckermann
Research and Technology Center Westcoast (FTZ)
Hafentoern, D-25761 Buesum, Germany
E-Mail recker@ftz-west.uni-kiel.de
Tel. +49-(0)4834-604-204 or -212 or -261
Fax. +49-(0)4834-604-299